|
The customer contacted a siemens customer care center (ccc).One quality control (qc) result on (b)(6) 2021 recovered high and out of range.Additional qc all recovered in range, and the calibration curves for the two reagent lots did not show any abnormalities.The samples were centrifuged correctly.No system abnormalities were identified.Discordant results were obtained using both the polypropylene and polystyrene sample tubes.There are no specifications or statements in the albumin's instructions for use nor in the bn prospec system instruction manual regarding the use of polystyrene tubes or polypropylene tubes.The cause of the event is unknown.The system is performing according to specifications.No further evaluation of this device is required.Mdrs 9610806-2021-00055 and 9610806-2021-00056 were filed for discordant results obtained on (b)(6) 2021 and (b)(6) 2021, respectively.
|
|
Three discordant, falsely elevated albumin results were obtained on a patient cerebrospinal fluid (csf) sample run in a polypropylene tube on a bn prospec system using n antiserum to albumin reagent.The discordant results were not reported to the physician(s).The sample was repeated for albumin csf in a polystyrene tube on the same system with the same reagent, recovering lower.The following day, the same sample was repeated for albumin csf in a polypropylene tube on the same bn prospec system using a new lot of n antiserum to albumin reagent, recovering falsely low.The sample was then repeated two additional times for albumin csf in a polystyrene tube on the same system with the same reagent, recovering higher.The results obtained in the polystyrene tube were considered correct.Two discordant, falsely low albumin results and one discordant, falsely elevated albumin result were obtained on a second patient csf sample run in a polypropylene tube on a bn prospec system using n antiserum to albumin reagent.The discordant results were not reported to the physician(s).The sample was repeated three times for albumin csf in a polystyrene tube on the same system with the same reagent, once recovering falsely high and twice recovering acceptably.The following day, the same sample was repeated for albumin csf in a polypropylene tube on the same bn prospec system using a new lot of n antiserum to albumin reagent, recovering falsely low.The sample was then repeated two additional times for albumin csf in a polystyrene tube on the same system with the same reagent, recovering higher.The lower albumin results obtained in the polystyrene tube were considered correct.A third patient csf sample was run two times for albumin in a polypropylene tube on a bn prospec system using n antiserum to albumin reagent.The following day, the same sample was repeated for albumin csf in a polypropylene tube on the same bn prospec system using a new lot of n antiserum to albumin reagent, recovering higher.The results were not reported to the physician(s).It is unknown what the correct albumin result was for this patient.Four discordant, falsely elevated albumin results were obtained on a fourth patient cerebrospinal fluid (csf) sample run in a polypropylene tube on a bn prospec system using n antiserum to albumin reagent.The discordant results were not reported to the physician(s).The sample was repeated four times for albumin csf in a polystyrene tube on the same system with the same reagent, recovering lower.The results obtained in the polystyrene tube were considered correct.There are no known reports of patient intervention or adverse health consequences due to the discordant albumin csf results.
|