The device was received at the local service facility for investigation.The technical service report is attached (see comm log), and indicates: the device is normal.Probable root cause: cables; connectors; digital board; power supply; filter/fuse; ac inlet board; main board; transition board; fiber board; intermittence between transition board and ch connector; software; scope (see rsk10356 output "visualization/image"); light source (see rsk10975 output "white light"); camera head (sensor, sensor cable); coupler (dust on optics, scratched optics, broken/shifted lenses, focusing mechanism, endoclamp, zoom ring); 1488 & 1588 extension cable; intermittence while using rf devices (ex: valleylab); problem toggling light source or from camera head; connected monitors; leaking; use error; poor grounding (e.G."finger" grounding tab connecting front and rear enclosure.Camera head connection from cable to transition board.); electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge; manufacturing/ service nonconformity.The reported failure mode will be monitored for future reoccurrence.
|