Model Number 328440 |
Device Problem
Volume Accuracy Problem (1675)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/19/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
|
Event Description
|
It was reported that syringe 0.3ml 31ga 8mm halfunit 10bg 500 was missing scale markings.This occurred on 6 occasions.The following information was provided by the initial reporter: the customer complained that some of the scale on the syringe is missing.
|
|
Manufacturer Narrative
|
H6: investigation summary customer returned (6) loose 3/10cc, 8mm syringes.Customer states that some of the scale on the syringe is missing.All returned syringes were examined and all exhibited missing scale markings from the 0-5 unit markings.A review of the device history record was completed for batch# 0293915.All inspections and challenges were performed per the applicable operations qc specifications.There was one (1) notification noted that did not pertain to the complaint.Confirmed: bd was able to duplicate or confirm the customer¿s indicated failure complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h10.
|
|
Event Description
|
It was reported that syringe 0.3ml 31ga 8mm halfunit 10bg 500 was missing scale markings.This occurred on 6 occasions.The following information was provided by the initial reporter: the customer complained that some of the scale on the syringe is missing.
|
|