| Model Number |
PSCST30 |
| Medical Device Problem Codes |
Crack (1135); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Weight and ethnicity: unknown, asked but unavailable.Implant date (2021 reports): if implanted, give date: not applicable as this is not an implantable device. explant date (2021 reports): if explanted, give date: not applicable as this is not an implantable device. email address: unknown, information not provided.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the cartridge cracked while inserting the intraocular lens (iol) and this ripped the lens.It was the cartridge part that is between the body of the cartridge that the lens is loaded into and the opening in the tip that the lens travels through which cracked.The lens was fully inserted into patient's operative eye, but was removed and replaced during the same procedure.Another johnson & johnson lens (same model and diopter) was implanted as a replacement.There was no injury to the patient and no medical/ surgical interventions were required.Patient was doing fine post-op.No further information was available.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9: device available for evaluation: yes section d9: returned to manufacturer on: 8/25/2021 section h3: device evaluated by manufacturer: yes device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification of the cartridge tip revealed viscoelastic residue inside the cartridge tip.The cartridge tip presented with stress marks, which is consistent for a cartridge that was used for implantation.Visual inspection under magnification of the complaint lens revealed viscoelastic residue on the optic body and haptics as well the lens being cut into three pieces.Based on the return condition of the lens no further product evaluation could be performed.The complaint issues could not be confirmed; therefore, no product deficiency could be identified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.The search revealed no additional complaints from this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|