Model Number 5550-G-391
Medical Device Problem Codes Fracture (1260); Migration (4003)
Health Effect - Clinical Code Pain (1994)
Date of Event 07/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
The following was reported that: "3 sheared patellar blocks, migration of the patella into the femur, pain".Patient required revision surgery on the (b)(6) 2021.
 
Event or Problem Description
The following was reported that: "3 sheared patellar blocks, migration of the patella into the femur, pain".Patient required revision surgery on the (b)(6) 2021.
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture involving a triathlon patella was reported.The event was confirmed via provided photo.Method & results: product evaluation and results: the reported device was not returned however a photograph was provided for review.The photo presented a patella with three pegs fractured off.Clinician review: no medical records were received for review with a clinical nsultant.Product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient's knee was revised due to patella fracture and migration of the patella into the femur.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON SYMMETRIC X3 PATELLA
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key12336357
Report Number0002249697-2021-01424
Device Sequence Number16850365
Product Code MBH
UDI-Device Identifier07613327014785
UDI-Public07613327014785
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2011
Device Model Number5550-G-391
Device Catalogue Number5550-G-391
Device Lot NumberW822
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/26/2021
Supplement Date Received by Manufacturer01/13/2022
Initial Report FDA Received Date08/18/2021
Supplement Report FDA Received Date02/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/21/2006
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-