Model Number 383400
Medical Device Problem Codes Inaccurate Flow Rate (1249); Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code Feeding Problem (1850)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.  if the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.  as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
 
Event or Problem Description
The customer reported that the device was overfeeding.
 
Additional Manufacturer Narrative
An evaluation of the kangaroo pump was performed.The unit was tested based on an accuracy test using 125ml of water.The delivered rate was as expected.The unit was within tolerance specifications.The unit¿s software was reinstalled as a preventative measure.The battery was replaced due to being out of date.The gasket and tie were replaced due to opening the unit.The pip pcb was replaced due to corrosion.The serial number label was replaced due to be illegible.No further actions are required at this time.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO JOEYPUMP W/POLE CLMP
Common Device Name
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
CARDINAL HEALTH
15 hampshire street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
15 hampshire street
mansfield MA 02048
Manufacturer Contact
jill saraiva
15 hampshire street
mansfield, MA 02048
5086183640
MDR Report Key12336486
Report Number1282497-2021-10551
Device Sequence Number20005277
Product Code LZH
UDI-Device Identifier10884521006645
UDI-Public10884521006645
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/11/2021
Supplement Date Received by Manufacturer08/11/2021
Initial Report FDA Received Date08/18/2021
Supplement Report FDA Received Date11/06/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-