| Model Number |
383400 |
| Medical Device Problem Codes |
Inaccurate Flow Rate (1249); Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
Feeding Problem (1850)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident sample has been requested but to date has not been received for evaluation. if the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted. as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
|
| |
|
Event or Problem Description
|
|
The customer reported that the device was overfeeding.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation of the kangaroo pump was performed.The unit was tested based on an accuracy test using 125ml of water.The delivered rate was as expected.The unit was within tolerance specifications.The unit¿s software was reinstalled as a preventative measure.The battery was replaced due to being out of date.The gasket and tie were replaced due to opening the unit.The pip pcb was replaced due to corrosion.The serial number label was replaced due to be illegible.No further actions are required at this time.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.
|
| |
|