Model Number 96-0880
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/23/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Original operation was performed (b)(6) 2017.Tibia was left in the patient.Surgeon replaced the implants with another tc3 femur, sleeve and a bigger pressfit stem.The procedure took 3 ½ hours.Affected side: left.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned impacted device did not find any evidence of product malfunction or product error.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SIGMA REVISION IMPLANT : KNEE FEMORAL ACCESSORY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12337137
Report Number1818910-2021-18081
Device Sequence Number2826187
Product Code NJL
UDI-Device Identifier10603295234364
UDI-Public10603295234364
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number96-0880
Device Catalogue Number960880
Device Lot Number103904
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2021
Initial Date Received by Manufacturer 07/26/2021
Supplement Date Received by Manufacturer09/07/2021
10/19/2021
Initial Report FDA Received Date08/18/2021
Supplement Report FDA Received Date09/17/2021
10/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MBT REVISION CEM TIB TRAY SZ 4; MBT TRAY SLEEVE POR M/L 29MM; PFC SIGMA FEM POST AUG SZ4 4MM; PFC SIGMA FEM POST AUG SZ4 8MM; PFC*SIGMA DIS AUG 4MM,SZ4,LEFT; PFC*SIGMA DIS AUG 4MM,SZ4,LEFT; PFC*SIGMA TC3 FEM LEFT SZ4; PFC*SIGMA/OV/DOME PAT 3PEG,38; SIG FEM ADAP +2/-2 OFFSET BOLT; SIGMA FEM ADAPTER 5 DEGREE; TC3 RP TIBIAL INSERT S4,12.5; UNIVERSAL STEM 75X16MM FLUTED; UNIVERSAL STEM 75X18MM FLUTED; MBT REVISION CEM TIB TRAY SZ 4; MBT TRAY SLEEVE POR M/L 29MM; PFC SIGMA FEM POST AUG SZ4 4MM; PFC SIGMA FEM POST AUG SZ4 8MM; PFC*SIGMA DIS AUG 4MM,SZ4,LEFT; PFC*SIGMA DIS AUG 4MM,SZ4,LEFT; PFC*SIGMA TC3 FEM LEFT SZ4; PFC*SIGMA/OV/DOME PAT 3PEG,38; SIG FEM ADAP +2/-2 OFFSET BOLT; SIGMA FEM ADAPTER 5 DEGREE; TC3 RP TIBIAL INSERT S4,12.5; UNIVERSAL STEM 75X16MM FLUTED; UNIVERSAL STEM 75X18MM FLUTED
Outcome Attributed to Adverse Event Required Intervention;
-
-