| Model Number |
417-152 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Unintended Radiation Exposure (4565)
|
| Date of Event |
07/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for investigation.
|
| |
|
Event or Problem Description
|
|
A peripheral atherectomy procedure commenced to treat a lesion in the proximal superficial femoral artery.The physician chose to use a spectranetics turbo elite laser atherectomy catheter to treat the patient.During the procedure, it was reported that the housing of the catheter was broken; when they inserted the catheter they could see red light emitting through the middle of the catheter.The procedure was completed with a second turbo elite device with no reported patient harm.This report captures the turbo elite which was reported damaged during use and resulted in unintended radiation exposure with potential for harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to the manufacturer on 14 sept 2021.The device investigation was completed on 16 sept 2021.The device was returned and evaluated by a cross functional team.Wrinkles were seen on the device''s outer jacket, 45.75in from the distal tip.Dead fibers and a breach to the outer jacket were seen in the area of the wrinkles.8-9 dead fibers were seen at both the distal tip and the proximal coupler.Likely, the device''s outer jacket became caught on an unknown structure, causing the wrinkles in and a breach to the outer jacket.This failure has been determined to be use related.
|
| |
|