| Model Number |
8439-1 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Fracture (1260)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
08/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The draw rod had been inserted into the nail and was tightened appropriately by the surgeon during the case.According to the surgeon's narrative, he grabbed a t8 screwdriver to actuate the reduction device, but instead inadvertently continued actuating the draw rod.
|
| |
|
Event or Problem Description
|
|
The event included the fracture of the tip of a draw rod.The fracture tip was left inside the driving feature of the nail, which was implanted in the patient.
|
| |
|
Search Alerts/Recalls
|