MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS LEGEND PNEUMATIC MOTOR; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS LEGEND PNEUMATIC MOTOR; MOTOR, DRILL, PNEUMATIC Back to Search Results
Model Number PM200
Medical Device Problem Codes Overheating of Device (1437); Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Report confirmed.The device was tested and the max temperature was measured to be 141°f.The likely cause was identified as corroded bearings.It was also noted that the vanes were worn and the hose was damaged.The device user manual warnings section includes instructions that heavy side loads and/or long operating times may cause the device to overheat.If the device overheats, discontinue use and rest the device by using it intermittently, or wrap the device in a moist sterile towel.Use of an overheated device may cause injury to the patient or operator.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Repair request initiated for device with the report of air leak and overheating.It was reported that there was no patient involvement.Repair escalated to a product event due to reason for return.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGEND PNEUMATIC MOTOR
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
MDR Report Key12339971
Report Number1625507-2021-00281
Device Sequence Number2825669
Product Code HBB
UDI-Device Identifier00721902499883
UDI-Public00721902499883
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPM200
Device Catalogue NumberPM200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2021
Initial Report FDA Received Date08/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-