| Model Number |
PM200 |
| Medical Device Problem Codes |
Overheating of Device (1437); Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Report confirmed.The device was tested and the max temperature was measured to be 141°f.The likely cause was identified as corroded bearings.It was also noted that the vanes were worn and the hose was damaged.The device user manual warnings section includes instructions that heavy side loads and/or long operating times may cause the device to overheat.If the device overheats, discontinue use and rest the device by using it intermittently, or wrap the device in a moist sterile towel.Use of an overheated device may cause injury to the patient or operator.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Repair request initiated for device with the report of air leak and overheating.It was reported that there was no patient involvement.Repair escalated to a product event due to reason for return.
|
| |
|