C.R. BARD, INC. (BASD) -3006260740 SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
|
Back to Search Results |
|
Catalog Number 9770552 |
Device Problems
Inappropriate/Inadequate Shock/Stimulation (1574); No Apparent Adverse Event (3189)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been received by the manufacturer for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
|
Event Description
|
After the procedure, when removing the power plug, there was an explosion, according to the engineer when they lifted the protective case of the source, this one with a darkened stain (burned).They didn't turn on the device anymore for fear of greater damage.
|
|
Event Description
|
After the procedure, when removing the power plug, there was an explosion, according to the engineer when they lifted the protective case of the source, this one with a darkened stain (burned).They didn't turn on the device anymore for fear of greater damage.
|
|
Manufacturer Narrative
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of burn mark from power supply was unconfirmed.The power supply is working correctly.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.H3 other text : evaluation findings are in section h.11.
|
|
Search Alerts/Recalls
|
|
|