| Model Number |
ADM05020013P |
| Medical Device Problem Codes |
Difficult to Remove (1528); Detachment of Device or Device Component (2907); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: the inpact admiral device was returned to medtronic investigation lab for evaluation.The device was returned with a guidewire entrapped in the device and a detachment at the shaft of the complaint device with the detached portion of the shaft and balloon entrapped in a non-medtronic sheath.Multiple sites of deformation were evident to the shaft of the device.Balloon bunching was evident immediately distal to the distal tip of the sheath with some fluid evident in the balloon.Slight deformation was present to the distal tip consistent with use of the device.The sheath was cut, and the entrapped section of the device was removed.The balloon folds were open.Multiple dents were evident to the shaft.A 0.035inch guidewire was loaded via the distal tip but was not possible to advance through the device due to the detachment and damage.Device removal recreation was not possible due to entrapment of the device in the sheath.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Physician was attempting to use inpact admiral paclitaxel eluting balloon catheter during procedure to treat the superficial femoral artery (sfa).The balloon was inflated using an inflation device.There was no damaged noted to packaging.There was no issue noted when removing the device from hoop/tray.The device was prepped per ifu with no issues identified.Removal difficulties occurred.It was reported that there was difficulty removing balloon following inflation.There was no resistance encountered when advancing the device.There were no difficulties noted during balloon deflation.The balloon was fully deflated when removal attempt made.The sheath and inpact admiral were removed in tandem.No intervention required for removal.Device was safely removed from patient.There was no patient injury.
|
| |
|
Search Alerts/Recalls
|