| Model Number |
PXB35-08-27-135 |
| Medical Device Problem Codes |
Inflation Problem (1310); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Physician was attempting to use a visipro balloon expandable stent along with 6fr sheath during procedure to treat a none calcified plaque lesion in the left distal subclavian artery.The vessel was moderately tortuous.There was no damage noted to packaging.There was no issue noted when removing the device from hoop/tray.The device was prepped per ifu with no issues identified.It was reported that the stent deployed in unintended lesion site (inaccurate delivery).The lesion site had sufficient landing zone.The balloon inflated unevenly and caused watermelon seeding of stent.This caused stent to slide off the shaft and be deployed distal to landing zone.An evercross pta (7x30) balloon was used to pre dilate the lesion.There was no resistance when advancing the device.Stent was deployed but not in the correct landing zone.Abnormal deployment left stent hanging into aorta.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: the stent displacement occurred due to inflation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation device returned coiled in 3 biohazard bags in a pouch.Device was decontaminated with cidex opa solution soak and tergazyme soak.The device was returned with no stent on the balloon.The deployment system showed no damage, and the balloon is in a post-inflated state.Image review one image was received from the customer.The image shows a deployed stent in the vasculature and what appears to be a guide catheter and a guidewire.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|