SMITH & NEPHEW, INC. ACC 2.0MM SS CABLE W/CLAMP; CERCLAGE, FIXATION Back to Search Results
Model Number 71340008
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a thr surgery, three (3) accord 2.0mm stainless steel cable with clamp were "fraying" before full tension was applied.Surgery was performed, after a non-significant delay, with a back-up device.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but are not limited to, excessive forces applied to implant or surgical technique.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACC 2.0MM SS CABLE W/CLAMP
Common Device Name
CERCLAGE, FIXATION
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12342860
Report Number1020279-2021-06470
Device Sequence Number11576968
Product Code JDQ
UDI-Device Identifier03596010525291
UDI-Public03596010525291
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K031162
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/11/2024
Device Model Number71340008
Device Catalogue Number71340008
Device Lot Number13MSM0119
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/05/2021
Supplement Date Received by Manufacturer10/11/2021
Initial Report FDA Received Date08/19/2021
Supplement Report FDA Received Date10/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-