Model Number RBYPOD8-A |
Device Problems
Fluid/Blood Leak (1250); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/23/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Event Description
|
The patient was undergoing a coil embolization procedure in the splenic artery using pod coils and a non-penumbra microcatheter.During the procedure, a pod coil had difficulty passing through the microcatheter, and the physician experienced resistance.Subsequently, the pod coil would not stay in the splenic artery.After multiple attempts to push the pod coil out from the distal tip of the microcatheter with the pod coil going in and out of the vessel, the pod coil was removed.The procedure was completed using a new pod coil and the same microcatheter.There was no report of an adverse effect to the patient.
|
|