| Catalog Number |
0375628000 |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foreign material in the sterile packaging.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: a foreign object was spotted inside the packaging of the sterile shaverblade.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be an improper cleaning process, qc incoming, and in-process inspection.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foreign material in the sterile packaging.
|
| |
|