| Model Number |
12360500 |
| Medical Device Problem Code |
Misassembled During Installation (4049)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product was returned as the explanted device was disposed of at the facility.No radiographs were provided confirming the device failure.The patient is recovering well post revision.Review of the provided information suggests the root cause of the issue is user error during installation as it was admitted that the device was misassembled during installation.No additional investigation can be completed."potential adverse events and complications.As with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries include.Potential risks identified with the use of this system, which may require additional surgery, include:.Bending, fracture or loosening of implant component(s).Loss of fixation." ".Warnings, cautions and precautions.The implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.".
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021 a patient underwent a posterior fixation procedure from t10-l5.In july on a unknown date during a routine post op follow up a radiograph revealed that the that the rod was installed in reverse against the curvature.According to information received, the bent rod was almost straight and so the screws were not loose and the patient had no pain.On (b)(6) 2021 a revision occurred where the device was replaced with a new rod.The revision operation was successfully completed and the patient is recovering well.
|
| |
|
Search Alerts/Recalls
|