MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT25042-4
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that signal loss over one hour occurred.The product was evaluated.An external/exterior visual inspection was performed and passed.Voltage check was performed and passed.Pairing test/download was performed and passed.A review of the bin file was performed and signal loss was not found within the investigation window.The allegation was not confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that signal loss over one hour occurred.The product was evaluated.An external/exterior visual inspection was performed and passed.Voltage check was performed and passed.Pairing test/download was performed and passed.A review of the bin file was performed and signal loss was not found within the investigation window.The allegation was not confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
MDR Report Key12345678
Report Number3004753838-2021-143880
Device Sequence Number4406955
Product Code QBJ
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
DEN170088
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/14/2021
Device Model NumberMT25042-4
Device Catalogue NumberSTT-OM-001
Device Lot Number7283963
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/25/2021
Supplement Date Received by Manufacturer08/23/2021
Initial Report FDA Received Date08/19/2021
Supplement Report FDA Received Date08/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age79 YR
Patient Weight79
-
-