• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number INX500S15
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Codes Unspecified Respiratory Problem (4464); Unspecified Gastrointestinal Problem (4491); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2021
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging an issue related to the device's sound abatement foam.There was no patient harm or injury.The manufacturer's investigation in on-going.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Event or Problem Description
The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.
 
Additional Manufacturer Narrative
The manufacturer previously reported an allegation of an issue related to sound abatement foam.This action was reported to fda per 21 cfr part 806.The device will be corrected per res 88058.
 
Additional Manufacturer Narrative
The manufacturer previously received information alleging an issue related to a cpap device's sound abatement foam.The customer had alleged to have black powder blowing out of the device.There was no report of serious patient harm or injury.In the initial report, the patient's allegation of feeling unwell with her stomach and nose was inadvertently omitted.There was no report of medical intervention.The manufacturer has corrected the health effect clinical code and health effect impact code in this report.The device was returned to a third-party service center for investigation.During the evaluation of the device, the third-party service center visually inspected the device and did not note any finding of sound abatement foam degradation.The device did not pass final testing.The device was scrapped.The manufacturer has updated h6 in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7243512041
MDR Report Key12347163
Report Number2518422-2021-03354
Device Sequence Number14703063
Product Code BZD
UDI-Device Identifier00606959025875
UDI-Public00606959025875
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberINX500S15
Device Catalogue NumberINX500S15
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/11/2021
Supplement Date Received by Manufacturer08/16/2021
08/26/2022
Initial Report FDA Received Date08/20/2021
Supplement Report FDA Received Date09/03/2021
10/20/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/03/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 88058
Patient Sequence Number1
Patient SexUnknown
-
-