MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Codes Positioning Failure (1158); Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: associated product tibia cemented 5 degree stemmed right size e item# 42532007102, lot# 64721507.Report source: (b)(6).Customer has indicated that the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that during a knee procedure, the surgeon failed to insert an articular surface to the tibial component.Several attempts were made and he confirmed that there was no cement or third particles between the articular surface and tibial component.Subsequently, a backup was used to complete the procedure.Attempts to obtain additional information have been made; however, no more information is available.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Visual inspection of the returned articular surface found that the dovetail feature was flared out.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key12348470
Report Number3007963827-2021-00176
Device Sequence Number5190432
Product Code MBH
UDI-Device Identifier00889024243750
UDI-Public(01)00889024243750(17)250131(10)64549154
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K121771
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model NumberN/A
Device Catalogue Number42522400710
Device Lot Number64549154
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/28/2021
Supplement Date Received by Manufacturer09/22/2021
Initial Report FDA Received Date08/20/2021
Supplement Report FDA Received Date09/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
-
-