Catalog Number 511100200060
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2021
Event Type  malfunction  
Manufacturer Narrative
The device evaluation was completed.A visual inspection was performed which observed two pins from one of the rings were not aligned with their mating holes on the opposing coupler ring.If the rings are not properly aligned, the coupler will not function as intended.Functional testing was not performed for this complaint.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During evaluation of a 2.0mm coupler, the rings were found to be misaligned; two pins from one of the rings were not aligned with their mating holes on the opposing coupler ring.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COUPLER
Type of Device
DEVICE, ANASTOMOTIC, MICROVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
SYNOVIS SURGICAL INNOVATIONS
2575 university ave. w
saint paul MN 55114
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key12348946
MDR Text Key267506549
Report Number1416980-2021-05132
Device Sequence Number1
Product Code MVR
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K861985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511100200060
Device Lot NumberSP20D30-1447752
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2021
Initial Date Manufacturer Received 08/06/2021
Initial Date FDA Received08/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-