Catalog Number 511100200060 |
Device Problem
Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/21/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device evaluation was completed.A visual inspection was performed which observed two pins from one of the rings were not aligned with their mating holes on the opposing coupler ring.If the rings are not properly aligned, the coupler will not function as intended.Functional testing was not performed for this complaint.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
During evaluation of a 2.0mm coupler, the rings were found to be misaligned; two pins from one of the rings were not aligned with their mating holes on the opposing coupler ring.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|