• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. ETHLN BLK 12IN 10-0 D/A CS160-6; SUTURE, NONABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. ETHLN BLK 12IN 10-0 D/A CS160-6; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Model Number 9000G
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: it was reported to me that 4 total needles disconnected from the sutures.It was more than one procedure but not specified how many.The nurse was not given that information.The lot # is: rcmazk.I have 8 each from the box to return.Please send return shipping instructions.The single complaint was reported with multiple events.There are no additional details regarding the additional patient events.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.It was reported that ¿it was more than one procedure but not specified how many.The nurse was not given that information¿ were these events previously reported to ethicon? if yes provide complaint file number for each procedure involved.Please include the following information: product code, lot number, complete event details, number of devices that had suture pull off issue during each procedure, device availability.Note: events reported via 2210968-2021-07556, 2210968-2021-07557 and 2210968-2021-07558.
 
Event or Problem Description
It was reported that a patient underwent an eye procedure on (b)(6) 2021 and suture was used.During the procedure, the suture got detached from the needle.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Date sent to the fda: 09/17/2021.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d4, d9, h4 investigation summary : an opened box with eight packets of product code 9000g was returned to ethicon inc for analysis with the packaging closed.Upon initial inspection, of the samples, no external damages were observed on the packet.In order to evaluate the conditions of the returned samples, the packets were opened, and no defects were detected.The swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed, and the pull force meets the requirements.The device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Device history lot :"a manufacturing record evaluation was performed for the finished device lot number rcmazk and no non-conformances related to the reported complaint condition were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHLN BLK 12IN 10-0 D/A CS160-6
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
MDR Report Key12348987
Report Number2210968-2021-07559
Device Sequence Number6026201
Product Code GAR
UDI-Device Identifier10705031022300
UDI-Public10705031022300
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model Number9000G
Device Catalogue Number9000G
Device Lot NumberRCMAZK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2021
Initial Date Received by Manufacturer 07/23/2021
Supplement Date Received by Manufacturer08/24/2021
Initial Report FDA Received Date08/20/2021
Supplement Report FDA Received Date09/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-