A user facility biomedical technician (biomed) reported that a 2008t machine was failing tmp tests.
Biomed is still troubleshooting the reported issue and stated the machine is still out of service.
The check valve, the dialysate pressure transducer #9, along with a few other parts have been replaced and the venous pressure was calibrated but none of those resolved the reported issue.
The tech explained they saw, on top of the balancing chamber, that a valve (possibly a solenoid valve) had melted.
Upon follow up, biomed did not observe any burning smell, smoke, spark, flame, arcing, or any other visible heat or electrical damage related to the melting found on the valve.
Biomed stated that the machine has approximately 13,000 hours and that the valve is the original fresenius part on the machine.
Biomed stated that the entire balance chamber was replaced, however the machine is still failing tmp tests for hi and low, and that the machine is not back in service.
Additionally, biomed confirmed that there was no damage observed on any other components.
Biomed confirmed that the machine has had not problems with passing electrical leakage tests and is plugged into a hospital grade ground fault circuit interrupter (gfci) outlet.
Biomed confirmed that a patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.
Biomed confirmed that the balance chamber and valve were discarded and are not available for return to the manufacturer for physical evaluation.
|