• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH CONNECT SFTY SCREW SPIKE NS; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH CONNECT SFTY SCREW SPIKE NS; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 77000FD
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.  if the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.  as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
 
Event or Problem Description
Customer reports: there is a leak at the junction of the spike.
 
Additional Manufacturer Narrative
The device history record (dhr) shows evidence that the product was released according to all established procedures and qa documentation.Three used samples and a box consisting of 22 unused samples were received at the manufacturing site for evaluation.A visual and functional inspection was performed on the returned samples and confirmed that the device leaked between the cross spike and detached cross spike from the line.The root cause and action plan will be documented through a formal corrective/preventative action (capa) which has been initiated to address the reported condition to mitigate any further occurrences.This action is currently ongoing.This complaint will be closed with no further action and will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECT SFTY SCREW SPIKE NS
Common Device Name
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
CARDINAL HEALTH
15 hampshire street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
15 hampshire street
mansfield MA 02048
Manufacturer Contact
jill saraiva
15 hampshire street
mansfield, MA 02048
5086183640
MDR Report Key12350696
Report Number1282497-2021-10558
Device Sequence Number14473532
Product Code LZH
UDI-Device Identifier10884521190818
UDI-Public10884521190818
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number77000FD
Device Catalogue Number77000FD
Device Lot Number192030034
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/18/2021
Supplement Date Received by Manufacturer08/18/2021
Initial Report FDA Received Date08/20/2021
Supplement Report FDA Received Date01/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-