• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PDSII VIO 27IN 4-0 S/A SH; SUTURE, SURGICAL, ABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PDSII VIO 27IN 4-0 S/A SH; SUTURE, SURGICAL, ABSORBABLE Back to Search Results
Model Number Z315H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4) attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: how many sutures are part of this complaint (please note there are 36 packets of sutures in a box) there were 3 sutures used in a row.Are the sutures that are part of this complaint available for collection? if so how many? are they contaminated? customer is sending remaining stock back a manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Photo investigation summary: the product was not returned to ethicon inc for evaluation.Visual inspection was conducted on the picture received.Visual analysis of the picture received determined that an opened box and an unopened sample of product code z315h could be observed.The condition reported could not be observed in the picture.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached as to what caused the reported complaint since the sample was not returned for analysis.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Related medwatch reports - 2210968-2021-07572 and 2210968-2021-07573.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date in 2021 and suture was used.The suture broke quite easily and was not normal quality.There were no adverse patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 09/23/2021.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d9.Investigation summary: an opened box with twenty-four packets of product code z315 were returned to ethicon inc for analysis with the packaging closed.Upon initial inspection of the samples, no external damages were observed on the packets.In order to evaluate the conditions of the returned samples, thirteen packets were opened, and no defects were detected.The swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along of the strand to detect any issue related to damaged or breakage sutures and no defects were observed during evaluation.Functional test was performed, and the tensile strength result were above the minimum requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDSII VIO 27IN 4-0 S/A SH
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
MDR Report Key12350737
Report Number2210968-2021-07571
Device Sequence Number1180420
Product Code NEW
UDI-Device Identifier10705031060456
UDI-Public10705031060456
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
N18331
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2025
Device Model NumberZ315H
Device Catalogue NumberZ315H
Device Lot NumberQPMEUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2021
Initial Date Received by Manufacturer 07/23/2021
Supplement Date Received by Manufacturer08/27/2021
Initial Report FDA Received Date08/20/2021
Supplement Report FDA Received Date09/23/2021
Patient Sequence Number1
-
-