| Model Number |
IPN914919 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#: (b)(4).
|
| |
|
Event or Problem Description
|
|
The user was inserting an emergent catheter in the icu.During insertion, just prior to passing the wire, the user noticed a well-em bedded hair within the guidewire and "acted appropriately".No patient harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The customer returned one guide wire assembly for evaluation.Visual examination did not reveal any defects or anomalies.No foreign materials were detected in or on the exterior of the guide wire.A device history record review was performed with no relevant findings to suggest a manufacturing related issue.The ifu provided with this kit cautions the user, "do not use if package has been previously opened or damaged." the customer report of a foreign material in the guide wire could not be confirmed by investigation of the returned sample.No defects or anomalies were detected with the returned guide wire.A device history record review was performed with no relevant findings to suggest a manufacturing related issue.No problem was found with the returned sample.Teleflex will continue to monitor and trend for complaints of this nature.
|
| |
|
Event or Problem Description
|
|
The user was inserting an emergent catheter in the icu.During insertion, just prior to passing the wire, the user noticed a well-em bedded hair within the guidewire and "acted appropriately".No patient harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Additional information received: the customer has provided a photo for evaluation.Visual inspection of the customer photo confirmed a hair was trapped in the guide wire assembly between the straight tubing and the thumb guide.The customer report of a foreign material in the guide wire assembly was confirmed by the returned customer photo.A device history record review was performed with no relevant findings.Based on the customer description and the customer photo, manufacturing likely caused or contributed to this event.A non-conformance was initiated to further investigate this issue.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
The user was inserting an emergent catheter in the icu.During insertion, just prior to passing the wire, the user noticed a well-em bedded hair within the guidewire and "acted appropriately".No patient harm reported.
|
| |
|