• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL NEEDLE 22X1-1/2 RB; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL NEEDLE 22X1-1/2 RB; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 305156
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used as a default.Investigation summary: a device history record review was completed by our quality engineer team for provided material number 305156 and lot number 0363742.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that needle 22x1-1/2 rb leaked.The following information was provided by the initial reporter: "it was reported: spontaneous: patient reports reconstitution syringe leaking/failing over the last few shipments she has received.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEEDLE 22X1-1/2 RB
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
MDR Report Key12351697
Report Number1911916-2021-00866
Device Sequence Number17425853
Product Code FMI
UDI-Device Identifier30382903051565
UDI-Public30382903051565
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K021475
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/31/2026
Device Model Number305156
Device Catalogue Number305156
Device Lot Number0363742
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2021
Initial Report FDA Received Date08/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-