| Model Number |
EG29-I10C |
| Medical Device Problem Codes |
Loss of or Failure to Bond (1068); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: as a result of investigating for the returned scope, we confirmed that a breakage of kapton tape wound around cmos unit inside the scope distal end.It caused that the electrical safety test in emea to fail.This device is classified as import for export, therefore 510k is not applicable.Model eg29i10c us is available in the usa with a 510k number k190805.This report is being filed as part of the pentax backlog management plan.
|
| |
|
Event or Problem Description
|
|
This even occurred during inspection.There was no report of patient harm.Electrical safety test failed, found during the incoming inspection by varimed.
|
| |
|