| Model Number |
EG29-I10C |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Optical Distortion (3000)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This event occurred at the time of during inspection.There was no report of patient harm.Iso failed ; misty image.
|
| |
|
Additional Manufacturer Narrative
|
|
This device is classified as import for export, therefore 510k is not applicable.Model eg29-i10c is available in the usa with a 510k number k190805.Evaluation summary we checked the returned product and confirmed that we confirmed that a breakage of kapton tape wound around cmos unit inside the scope distal end.It caused that the electrical safety test in emea to fail.Regarding the misty image,it was caused due to a condensation of moisture in the ccd unit by changing the temperature outside of the endoscope.This report is being filed as part of the pentax backlog management plan.
|
| |
|