C.R. BARD, INC. (BASD) -3006260740 BARD BUTTON REPLACEMENT GASTROSTOMY DEVICE; FEEDING DEVICE Back to Search Results
Model Number 000284
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a video was provided for review.The investigation of the reported event is currently underway.(expiry date: 01/2023).
 
Event or Problem Description
It was reported that post feeding device implant, the device was allegedly leaking profusely.It was further reported that the device was replaced.There was no reported patient injury.
 
Event or Problem Description
It was reported that sometime post feeding device implant, the device was allegedly leaking profusely.It was further reported that the device was replaced.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the sample was not returned for evaluation.One electronic video was provided for review.The investigation is confirmed for the leak issue as a leak was found between the adaptor port and closure strap area.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD BUTTON REPLACEMENT GASTROSTOMY DEVICE
Common Device Name
FEEDING DEVICE
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key12352008
Report Number3006260740-2021-03450
Device Sequence Number17127733
Product Code KGC
UDI-Device Identifier10801741087872
UDI-Public(01)10801741087872
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K904779
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number000284
Device Catalogue Number000284
Device Lot NumberHUEP0985
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/26/2021
Supplement Date Received by Manufacturer09/01/2021
Initial Report FDA Received Date08/22/2021
Supplement Report FDA Received Date09/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-