OLYMPUS MEDICAL SYSTEMS CORP. HALOGEN LIGHT SOURCE Back to Search Results
Model Number CLH-SC
Device Problem Energy Output Problem (1431)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus service operation repair center (sorc).Sorc checked the subject device and found that the reported phenomenon could be duplicated.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed by the user that during an unspecified timing, the device didn't work and the lamp didn't light on even though it was replaced.There was no report of patient injury associated with the event.The event date was unknown.The device is an olympus asset.
 
Manufacturer Narrative
This is a supplemental report to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.The manufacturing record could not be reviewed because the subject device was manufactured more than 15 years ago.It was considered that there was a high possibility that the reported phenomenon occurred due to the influence that the temperature fuse broke.The service life is 6 years.Since the device was manufactured for more than 20 years ago and it was used for a long time, it was considered that there was a high possibility that the temperature fuse broke and the reported phenomenon occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LIGHT SOURCE
Type of Device
HALOGEN LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12352207
MDR Text Key267627836
Report Number8010047-2021-10583
Device Sequence Number1
Product Code GCT
Combination Product (y/n)N
PMA/PMN Number
K993041
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberCLH-SC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2021
Was the Report Sent to FDA? No
Date Manufacturer Received10/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-