| Medical Device Problem Codes |
Overheating of Device (1437); Electrical Power Problem (2925)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined that the device was not powering on from wall power and that the power inlet module and cord were burnt.The power inlet module was replaced and resolved the reported issue.It was noted during testing that the unit had low suction and the inline filter and vacuum regulator were replaced.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported this unit will not power on via wall power.Found it will power on via the evacuation unit.The event timing was during cleaning.Investigation found the power inlet module was burnt and the cord had heat damage.No adverse event was reported as a result of this malfunction.
|
| |
|