MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AESTIVA 7900; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AESTIVA 7900; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2021
Event Type  malfunction  
Manufacturer Narrative
No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The speaker was replaced to resolve the reported issue.
 
Event Description
The hospital reported a malfunction resulting in the loss of audible alarms.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESTIVA 7900
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
john szalinski
3030 ohmeda drive
madison, WI 53718
MDR Report Key12353190
MDR Text Key268917625
Report Number2112667-2021-01937
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 08/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/26/2021
Initial Date FDA Received08/23/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-