| Model Number |
82090-01 |
| Medical Device Problem Codes |
Mechanical Jam (2983); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Visual and functional analysis was performed on the returned device.The reported deployment failure was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar complaints.The investigation was unable to determine a definitive cause for the reported difficulties.It may be possible that the distal outer sheath was bent or entrapped in the anatomy causing resistance with the thumbwheel and preventing deployment; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat an internal carotid artery that was 90% stenosed.An xact stent was advanced to the lesion and deployment initiated.However, it was noted that the release knob of the device could not be turned normally and the stent completely failed to deploy.A new xact stent was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.Additional per device analysis confirmed that [the distal end of the stent implant was partially expanded over the tip for a length of 6mm].No additional information was provided.
|
| |
|