• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XACT CAROTID STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XACT CAROTID STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Model Number 82090-01
Medical Device Problem Codes Mechanical Jam (2983); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Visual and functional analysis was performed on the returned device.The reported deployment failure was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar complaints.The investigation was unable to determine a definitive cause for the reported difficulties.It may be possible that the distal outer sheath was bent or entrapped in the anatomy causing resistance with the thumbwheel and preventing deployment; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat an internal carotid artery that was 90% stenosed.An xact stent was advanced to the lesion and deployment initiated.However, it was noted that the release knob of the device could not be turned normally and the stent completely failed to deploy.A new xact stent was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.Additional per device analysis confirmed that [the distal end of the stent implant was partially expanded over the tip for a length of 6mm].No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XACT CAROTID STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key12353852
Report Number2024168-2021-07382
Device Sequence Number3516591
Product Code NIM
UDI-Device Identifier08717648010262
UDI-Public08717648010262
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P040038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model Number82090-01
Device Catalogue Number82090-01
Device Lot Number0041561
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/11/2021
Initial Report FDA Received Date08/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-