MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. OES CYSTONEPHROFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. OES CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5
Device Problems Break (1069); Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2021
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.The evaluation confirmed the user report.The bending section sheath broke off and moisture invaded the scope.The moisture invasion caused a stain on the image.This event is under investigation.A supplemental report will be submitted upon receiving additional information.
 
Event Description
The customer reported the cystonephrofiberscope had exposed metal where the rubber was peeling off of the bending section.Per the customer, this was found during reprocessing.There was no patient involvement or impact to patient care reported for this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, it is likely that the rubber a was damaged due to an external force applied to the curved part, leading to the exposure of the metal part.This issue is addressed in the instructions for use (ifu): "important information please read before use warnings and cautions follow the warnings and cautions given below when handling this instrument.This information is to be supplemented by the warnings and cautions given in each chapter.¿ never perform angulation control, suction control or insertion/withdrawal of the endoscope¿s insertion tube without viewing the endoscopic image.Patient injury,bleeding and/or perforation can result.¿ never insert or withdraw the endoscope¿s insertion tube while the up/down angulation is locked.Patient injury and/or equipment damage can result." olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12353930
MDR Text Key267676103
Report Number8010047-2021-10615
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170339417
UDI-Public04953170339417
Combination Product (y/n)N
PMA/PMN Number
K032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 09/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2021
Was the Report Sent to FDA? No
Date Manufacturer Received09/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-