• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS INC DREAMSTATION AUTO CPAP W/HUMID/HT, INTL; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS INC DREAMSTATION AUTO CPAP W/HUMID/HT, INTL; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number INX500T15
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Codes Dyspnea (1816); Liver Damage/Dysfunction (1954); Chronic Obstructive Pulmonary Disease (COPD) (2237); Respiratory Failure (2484); Cancer (3262)
Date of Event 08/10/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The manufacturer received information alleging a continuous positive airway pressure (cpap) device's sound abatement foam became degraded and caused a patient to develop chronic obstructive pulmonary disease (copd).This issue was reported to the fda per 21 cfr 806.The device will be corrected per (b)(4).
 
Additional Manufacturer Narrative
The manufacturer previously reported an allegation of an issue related to sound abatement foam.  a correction to b5 was made and should be reported as:    the manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.The patient alleged to develop shortness of breath with chronic respiratory failure and chronic obstructive pulmonary disease (copd) and ruptured liver cancer.The patient was hospitalised in response to the event.The reported event of shortness of breath with chronic respiratory failure and chronic obstructive pulmonary disease (copd) and ruptured liver cancer and its reported severity was reviewed by the manufacture's clinical expert.This event is assessed as not related to the device in this case.Based on the available information, the manufacture concludes no further action is necessary.The device was returned to the manufacturer's product investigation laboratory for further investigation.  the device was evaluated.There was no mention of visual findings to the external part of the device.  the internal aspect of the device was inspected.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the manufacturer.There were no errors found.   the manufacturer concludes that they could not confirm the customer's allegation and there was no visible foam degradation.Section(s) b1, b2, has changed related to the complaint changing from the reported adverse event to a product problem.Section h1 has changed to reflect a malfunction.Section h6 health effect- clinical code was corrected (missed to capture all the codes in initial mdr) and section h6 health effect- impact code, type of investigation findings and investigation conclusions has been updated.
 
Additional Manufacturer Narrative
The manufacturer previously submitted mdr 2518422-2021-03066-1 with incorrect sections b1, b2, h1, h6.Corrections to previous mdr are made in this report as follows.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP W/HUMID/HT, INTL
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS INC
1001 murry ridge ln
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS INC
1001 murry ridge ln
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key12354648
Report Number2518422-2021-03066
Device Sequence Number3726381
Product Code BZD
UDI-Device Identifier00606959025899
UDI-Public00606959025899
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberINX500T15
Device Catalogue NumberINX500T15
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2021
Supplement Date Received by Manufacturer03/21/2023
03/21/2023
Initial Report FDA Received Date08/23/2021
Supplement Report FDA Received Date04/19/2023
06/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;