• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE; NEBULIZER (DIRECT PATIENT INTE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE; NEBULIZER (DIRECT PATIENT INTE Back to Search Results
Model Number IPN040620
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review could not be conducted since the lot number of the device was not provided.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event or Problem Description
It was reported that "the connection part between the nut adaptor and the flowmeter was so unstable that the adaptor got detached during use.Therefore, it was replaced with a new unit".No patient harm or injury reported.Patient condition reported as "fine".
 
Event or Problem Description
It was reported that "the connection part between the nut adaptor and the flowmeter was so unstable that the adaptor got detached during use.Therefore, it was replaced with a new unit".No patient harm or injury reported.Patient condition reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and it was observed that the puncture pin was missing.No other issues were observed.A puncture pin was obtained and oxygen entrainment testing was performed.During the setup it was observed that the assembly of the nut adaptor and the upper body was unstable.Even with that condition, the sample was able to be tested on oxygen entrainment testing, but the sample failed the test.A visual inspection was performed and wear was found on the internal tabs of the adaptor.Based on the investigation performed, the complaint is confirmed.Although the complaint is confirmed, there is not sufficient evidence to assure that this issue was originated during the manufacturing assembly or molding process.The wear on the internal tabs of the adaptor was most likely caused by the end user during the connection of the adaptor into the flowmeter that causes an unstable connection issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE
Common Device Name
NEBULIZER (DIRECT PATIENT INTE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key12355437
Report Number3004365956-2021-00262
Device Sequence Number4115976
Product Code CAF
UDI-Device Identifier04026704581290
UDI-Public04026704581290
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153010
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberIPN040620
Device Catalogue Number031-33J
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2021
Initial Date Received by Manufacturer 08/02/2021
Supplement Date Received by Manufacturer09/09/2021
Initial Report FDA Received Date08/23/2021
Supplement Report FDA Received Date09/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.; N/A.
-
-