| Model Number |
8780 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a healthcare provider (hcp) via a company representative (rep) regarding a patient receiving prialt (25 mcg/ml, 1.2 mcg/day) via an implantable pump for an unknown indication of use.It was reported that when the physician had picked up the anchor to implant, it was not attached as expected to the deployment tool.The anchor was hanging to the side with the green piece outside of the anchor tool.Instead of trying to utilize, an 8785 was opened and implanted.There were no known environmental/external factors that may have led or contributed to the issue.The issue was resolved at the time of this report.The patient's medical history included diabetes, no other known medical history.The patient was alive with no injury at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis confirmed that the handle of the anchor deployment tool was separated from the hypotube.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|