Model Number REACT-71
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 08/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that when the doctor went to pull the react 71 catheter out they noticed that the distal radiopaque tip had separated and was lodged in the anterior cerebral artery (aca).It was noted that the tip was stuck, but the catheter was not stuck inside in the guide catheter.No further medical/surgical interventions were required, and the tip remained in the aca. the patient did not experience any injury.The devices were prepared and flushed according to the instructions for use (ifu).The patient was undergoing a thrombectomy in the internal carotid artery/middle cerebral artery (ica/mca).The patient's vessel tortuosity was moderate.Ancillary devices include solitaire and trevo stent retrievers and a zoom 88 catheter.
 
Event or Problem Description
Additional information received reported that there were no issues that led to the marker band dislodgement.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REACT CATHETER
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key12356025
Report Number2029214-2021-01050
Device Sequence Number17023258
Product Code DQY
UDI-Device Identifier00847536031784
UDI-Public00847536031784
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K182097
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/26/2022
Device Model NumberREACT-71
Device Catalogue NumberREACT-71
Device Lot NumberB049175
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2021
Supplement Date Received by Manufacturer09/01/2021
Initial Report FDA Received Date08/23/2021
Supplement Report FDA Received Date09/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age59 YR
-
-