WRIGHTS LANE SYNTHES USA PRODUCTS LLC 12NM TORQ LIMITING WRENCH F/ DUAL-OP USS Back to Search Results
Model Number 03.602.042
Medical Device Problem Code Calibration Problem (2890)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reporter is a j&j employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that, during testing at service and repair, a torque limiting handle failed in calibration.No patient involvement.This report is for 1 12nm torque limiting wrench for dual-opening uss close.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: part # 03.602.042.Synthes lot # h484541-01.Supplier lot # h484541-01.Release to warehouse date: 10 may2019, 13 may2019.Supplier: (b)(4).No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Service and repair evaluation: during testing at service and repair, the repair technician reported a torque wrench failed low in calibration testing.Torque test failed low is the reason for repair.The cause of the issue is unknown.The item will be repaired per the work instructions.And will be returned to the customer upon completion of the service and repair process.Finalized service record will be archived in tungsten document management system.The evaluation was confirmed.The device was deemed serviceable and will be returned to the customer, no design or manufacturing issues were identified therefore it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12NM TORQ LIMITING WRENCH F/ DUAL-OP USS
Common Device Name
WRENCH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key12356196
Report Number2939274-2021-04815
Device Sequence Number3420951
Product Code HXC
UDI-Device Identifier10705034712277
UDI-Public(01)10705034712277
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.602.042
Device Catalogue Number03.602.042
Device Lot NumberH484541-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2021
Initial Date Received by Manufacturer 07/26/2021
Supplement Date Received by Manufacturer09/29/2021
Initial Report FDA Received Date08/23/2021
Supplement Report FDA Received Date09/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-