SYNTHES GMBH TFNA HELICAL BLADE 80MM STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
Catalog Number 04.038.280S |
Device Problem
Migration (4003)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2020, the patient underwent the open reduction internal fixation (orif) surgery for the trochanteric fracture of the femur with the trochanteric fixation nail-advanced (tfna) implants.After the surgery, over-sliding was observed in the x-ray during follow-up, and since there was no pain when walking in a wheelchair, the surgeon was following up the patient.But the x-ray on (b)(6) 2021, showed cut out, and the doctor judged that fusion of the fracture had been achieved, resulting in implant removal.After removal of the implant, the bone prosthetic material is filled into the space and the surgeon closed the incision.The surgery was completed successfully without any surgical delay.The surgeon commented that the cut-out was not caused by the products.No further information is available.This report is for one (1) tfna helical blade 80mm sterile.This is report 2 of 4 for complaint (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6: part number: 04.038m280sp.Lot number: 67p1831.Part manufacture date: 24-aug-2020 august 24, 2020.Part expiration date: n/a.Nonconformance noted: n/a.Dhr record review: a review of the device history record revealed no complaint related anomalies.The devices history record show this lot tfna helical blade 80mm product was processed through the normal manufacturing and inspection operations with no rework nor nonconformities noted.This lot was shipped out for final packaging.This lot met all dimensional, visual, and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.A review of the raw material device history record revealed this raw material lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|