• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION SD ALARIS SYSTEM; CARBON DIOXIDE GAS ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION SD ALARIS SYSTEM; CARBON DIOXIDE GAS ANALYZER Back to Search Results
Model Number 8300
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/03/2021
Event Type  malfunction  
Manufacturer Narrative
The actual date of event is unknown.A review of the complaint history was performed which did not confirm similar complaints with the same or related failure mode.A review of the device history record was performed from the date of manufacture to the date of product release for distribution which confirmed that this device was not involved in a production failure which correlates to the customer reported issue.A review of the service history record was performed which confirmed that this device was not involved in a service failure which correlates to the customer reported issue.Device evaluated by bd.
 
Event Description
It was reported that the device had alarm error codes/messages and etco2 board comm displayed error code 571.6241.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS SYSTEM
Type of Device
CARBON DIOXIDE GAS ANALYZER
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
brett wilko
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key12357579
MDR Text Key267825362
Report Number2016493-2021-58648
Device Sequence Number1
Product Code CCK
UDI-Device Identifier10885403830013
UDI-Public(01)10885403830013
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031741
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number8300
Device Catalogue Number8300 ALARIS ETCO2 MODULE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/03/2021
Initial Date FDA Received08/24/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-