Health effect impact code updated.The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A complaint history review concluded this was a repeat issue.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.Visual inspection found that the device was returned outside of it's original packaging.Two curved needles, one straight needle, and two suture loops were returned in the package.The returned straight needle has debris throughout.One of the suture loops is separated from the hub and there is debris on the shaft.The complaint was confirmed and the root cause was associated with device design.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A corrective action for this failure mode is in place.
|