• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. ON-Q PAIN RELIEF SYSTEM WITH SELECT-A-FLOW, 400 ML, 2-14 ML/HR; ELASTOMERIC - SAF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. ON-Q PAIN RELIEF SYSTEM WITH SELECT-A-FLOW, 400 ML, 2-14 ML/HR; ELASTOMERIC - SAF Back to Search Results
Model Number CB004
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 08/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record for lot 80400859 was reviewed and the product was produced according to product specifications.The actual complaint product was not returned for evaluation.Root cause could not be determined.All information reasonably known as of 23 aug 2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc.Complaint database and identified as complaint (b)(4).
 
Event or Problem Description
Fill volume: 500 ml.Flow rate: 8 ml/hr.Procedure: "achilles procedure." cathplace: unknown.Infusion start time: (b)(6) 2021 0920.Infusion stop time: (b)(6) 2021 0400.It was reported that "fast flow" event occurred.The patient called and stated the pump was "put in at 0745 after achilles procedure on friday (b)(6) [(b)(6) 2021] that was running at 8 ml/hr the entire time ran out on sunday morning at about 0400." pump infused in less than 2 full days.No side effects reported, but patient was in pain and was told the pump would run out on monday.Additional information received 03-aug-2021 indicated the infusion was started (b)(6) 2021 at 0920.The product was used in accordance with the instructions and facility conditions did not contribute to this event.The facility is not a new or inexperienced user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ON-Q PAIN RELIEF SYSTEM WITH SELECT-A-FLOW, 400 ML, 2-14 ML/HR
Common Device Name
ELASTOMERIC - SAF
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
MDR Report Key12361422
Report Number2026095-2021-00086
Device Sequence Number19786181
Product Code MEB
UDI-Device Identifier00193494134723
UDI-Public00193494134723
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K063530
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source consumer,health professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/24/2023
Device Model NumberCB004
Device Catalogue Number101347203
Device Lot Number80400859
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2021
Initial Report FDA Received Date08/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
0.25% MARCAINE.
Patient Age37 YR
Patient Weight89
-
-