| Model Number |
SOFTECHD+14.5 |
| Medical Device Problem Code |
Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/26/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A full audit of all batch documentation relating to the production of the device was performed.The audit concluded that all procedures in manufacturing and packaging of the device had been conducted correctly.Batch reconciliation was 100.0%.The visual inspection performed on the device determined that the lens was damaged due to being improperly handled.
|
| |
|
Event or Problem Description
|
|
Lenstec received an email stating " lens was implanted and physician noted there was no haptic.Removed and replaced with same type and diopter lens using an enlarged incision without patient issues.".
|
| |
|