• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - OUTFLOW GRAFT; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - OUTFLOW GRAFT; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1103
Medical Device Problem Codes Obstruction of Flow (2423); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Heart Failure/Congestive Heart Failure (4446)
Date of Event 01/08/2018
Type of Reportable Event Death
Additional Manufacturer Narrative
Investigation of this event is pending and a supplemental report will be sent upon its completion.Additional information will be requested regarding any further available additional details of the event, but it was not available at the time of this report.If additional information is received, the event will be updated and a supplemental report will be sent.This event was collected during a review of patient records with the healthcare facility.Additional products: product name: heartware ventricular assist system ¿ pump, model #: 1103 / catalog #: 1103 / expiration date: / serial#:(b)(4) / device available for evaluation?: no / labeled for single use?: no.(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient experienced outflow graft stenosis.The outflow graft stenosis was treated with an atrium covered stent.The patient subsequently expired from right ventricular failure.No further details were reported as part of the event.This event was collected during a review of patient records with the healthcare facility.
 
Additional Manufacturer Narrative
A supplemental report is being submitted for completion of the investigation and additional patient and device information.Product event summary: (b)(6) and the associated outflow graft (1001895862) were not returned for evaluation.Log file analysis could not be performed since log files covering the reported event date were not available for analysis.The reported event could not be confirmed due to insufficient evidence.Based on the available information, there is no evidence to indicate that a device malfunction or performance issue caused or contributed to the reported event.Per the instructions for use, this event is a known potential complication associated with the implantation of a vad.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to this event.This event was collected during a review of patient records with the healthcare facility.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - OUTFLOW GRAFT
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key12362451
Report Number3007042319-2021-06086
Device Sequence Number5156694
Product Code DSQ
UDI-Device Identifier00888707000222
UDI-Public00888707000222
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 10/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/31/2017
Device Model Number1103
Device Catalogue Number1103
Device Lot Number1001895862
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/30/2021
Supplement Date Received by Manufacturer09/20/2021
Initial Report FDA Received Date08/24/2021
Supplement Report FDA Received Date10/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1103; 1103
Outcome Attributed to Adverse Event Death;
Patient Age67 YR
-
-