• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 45 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 45 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 45-550
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 05/18/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture baker grade iv is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture.
 
Event or Problem Description
Healthcare professional reported left side "capsular contracture, baker grade iv".Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation:visual analysis of the returned device identified the weight was to the spec, a deformation, and a crease fold.Cloudy was observed after autoclave disinfection process.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.
 
Event or Problem Description
Healthcare professional reported left side "capsular contracture, baker grade iv".Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 45 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
MDR Report Key12362756
Report Number9617229-2021-49040
Device Sequence Number3735954
Product Code FTR
UDI-Device Identifier10888628002104
UDI-Public10888628002104
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2021
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2019
Device Model Number45-550
Device Catalogue Number45-550
Device Lot Number2687726
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2021
Initial Date Received by Manufacturer 07/27/2021
Supplement Date Received by Manufacturer08/30/2021
Initial Report FDA Received Date08/24/2021
Supplement Report FDA Received Date09/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age34 YR
-
-