MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number MERGE HEMODYNAMICS 10.1.2
Medical Device Problem Code No Device Output (1435)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2021
Type of Reportable Event Malfunction
Event or Problem Description
Merge hemo displays, measures, and records physiological data from a patient undergoing a cardiac catheterization procedure.The hemo system can visualize and capture vital sign values including ecg, impedance respiration, spo2 and pleth waveforms, invasive blood pressure, temperature, non-invasive blood pressure (nibp), thermodilution cardiac output and fractional flow reserve (ffr).The system can display and capture diagnostic quality 12 lead resting ecg to visualize arrhythmias, and st-segment changes.Some hemo systems have an option to measure and display side-stream end tidal carbon dioxide (etco2) along with apnea and respiration rates calculated from the etco2 waveform.The hemodynamic portion of the system is comprised of the patient data module (pdm) and the merge hemo monitor pc.The two units are connected via a serial interface.All vital parameters are acquired and calculated in the pdm.This data is then transmitted to the merge hemo monitor pc via the serial interface.All data can then be displayed on the merge hemo monitor pc.The merge hemo system is not intended to produce alarms for out-of-range conditions.Patient allergies and current medication information can be entered by the user and displayed by the system.If desired and using a third party database the hemo system can display drug to drug or drug to allergy interaction information.The system is intended for use in hospital cardiac catheterization laboratories and in pre-and post-procedure care areas in the hospital under the close supervision of qualified medical personnel.On (b)(6) 2021, a customer reported that a hemo ups (universal power supply) had caught fire and melted and the customer needed a replacement sent.Merge healthcare technical support confirmed that no patient was involved when this took place.Merge healthcare technical support has requested the ups in question to be sent back, and has sent a replacement as well as 2 power strips.The ups in question will be reviewed and a supplemental report will be filed when more information becomes available.This has the potential to delay patient treatment and/or diagnosis.There have been no reports of patient injury or harm as a result of this issue.Reference complaints (b)(4).
 
Additional Manufacturer Narrative
This supplemental report is submitted to the fda in accord with applicable regulations and as indicated by merge healthcare in the initial report submitted 8/24/2021.Inspection of the unit in question showed no evidence of burning of any kind.Merge healthcare support cleaned the batteries and ran the unit for 2 weeks without problems.No further action is required.Revised information contained in this supplemental report includes the following: g6 - indication that this is follow-up report 001, h2 - indication of additional information, h6 - investigation findings: 3221 no findings available, h6 - investigation conclusions: 4309 cause traced to environment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Common Device Name
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
brian bell
75 binney street
cambridge, MA 02142-1123
3123540189
MDR Report Key12362826
Report Number2183926-2021-00021
Device Sequence Number17096932
Product Code DQK
UDI-Device Identifier00842000100034
UDI-Public(01)00842000100034(10)10105380(11)170918
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K152864
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMERGE HEMODYNAMICS 10.1.2
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2021
Supplement Date Received by Manufacturer11/12/2021
Initial Report FDA Received Date08/24/2021
Supplement Report FDA Received Date11/12/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/18/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-