• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG LIGATURE CLIP 12 MAG.= 144 PCS.; LIGATION / VESSEL CLIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG LIGATURE CLIP 12 MAG.= 144 PCS.; LIGATION / VESSEL CLIPS Back to Search Results
Model Number PL572T
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported to aesculap ag that a ligature clip 12 mag.= 144 pcs (part # pl572t) was used during a procedure performed on (b)(6) 2021.According to the complainant, during the surgery, the magazine was deformed.The tip of the magazine was deformed (bent) and fell into the patient's body.The product has been recovered and there is no health hazard to the patient.The plate has come into contact with an obstacle or is damaged.The complaint device has not been received by the manufacturer for evaluation.No patient complications were reported as a result of the event.Although requested, additional information has not been made available.The adverse event is filed under aag reference (b)(4).
 
Additional Manufacturer Narrative
Updated: b5 intervention required; d9 no product return; h6 codes.Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
 
Event or Problem Description
Clarification: an additional medical intervention was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGATURE CLIP 12 MAG.= 144 PCS.
Common Device Name
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key12363122
Report Number9610612-2021-00583
Device Sequence Number3736586
Product Code FZP
UDI-Device Identifier04038653341856
UDI-Public4038653341856
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K081031
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/13/2025
Device Model NumberPL572T
Device Catalogue NumberPL572T
Device Lot Number52630778
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2021
Supplement Date Received by Manufacturer06/02/2022
Initial Report FDA Received Date08/24/2021
Supplement Report FDA Received Date06/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-