| Model Number |
Z9002 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Weight ¿ ethnicity: unknown, not provided.Date of event is unknown/not provided.If implanted; give date: n/a (not applicable).Lens was removed/replaced during the same procedure.If explanted; give date: n/a (not applicable).Lens was removed/replaced during the same procedure.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a tecnis monofocal intraocular lens (iol) was removed and replaced with a back up lens (same model/diopter) during the initial procedure due to the lens being torn.There was no harm to the patient and she has recovered from the surgery.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: no complaint lens was received, and therefore no product evaluation could be performed on the lens.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.Historical data analysis: a search found one complaint as part of this production order and was void.No escalations are required.Another complaint was found as also part of this production order however, there has been no investigation initiated for this complaint at this time, therefore, no product malfunction or deficiency could be identified, and no escalations are required.Conclusion: as a result of the investigation, there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|